ISO 8655, As Found and As Returned: Managing Pipette Calibration in GMP-Compliant Pharmaceutical Laboratories

Pipettes are among the most frequently used instruments in a pharmaceutical laboratory, and also among the most easily overlooked. ISO 8655 sets out the requirements for verifying their performance, while the as found and as returned values recorded on a calibration certificate tell two very different stories about the instrument. Understanding what each of these values means, and how they fit into the quality system, is what separates a formally calibrated pipette from one that genuinely supports the defensibility of analytical data.

Why Pipette Calibration is so Critical in GMP Laboratories

In pharmaceutical laboratories, the quality of analytical data depends on the ability to guarantee accurate, repeatable and traceable measurements over time. Pipettes are used daily in high-criticality activities such as standard preparation, dilutions, microbiological controls and sample preparation. Even apparently minimal volumetric variations can have a significant effect on final results, particularly in regulated environments where data repeatability is an essential requirement.

The main problem is that volumetric error is often not immediately perceptible to the operator. A pipette can continue to function apparently correctly while having lost accuracy or precision over time. This is precisely one of the most critical aspects in GMP laboratories: instruments that appear to be “working” may in fact compromise the reliability of the entire analytical process.

ISO 8655: The Normative Reference for Pipette Calibration

The ISO 8655 standard is the principal international reference for verifying the performance of pipettes and piston-operated volumetric apparatus.

The standard defines:

  • accuracy and precision requirements;
  • systematic and random error limits;
  • environmental test conditions;
  • gravimetric verification procedures;
  • instrument conformity criteria.

Compliance with ISO 8655 is not only a technical requirement, but also a fundamental tool for ensuring the robustness of the quality system and the defensibility of analytical data during audits and inspections.

As Found and As Returned: What They Really Mean

Among the most important elements in pipette calibration management are the as found and as returned values, which frequently appear on certificates but are not always correctly interpreted.

The as found value represents the actual condition of the pipette before technical intervention. It is the value that makes it possible to understand whether the instrument was operating within tolerance during its operational use. This aspect is particularly important in GMP laboratories because an out-of-tolerance condition detected at the as found stage may require a retrospective evaluation of previously generated analytical results.

The as returned value, by contrast, indicates the performance of the instrument after adjustment or maintenance and confirms that correct volumetric performance has been restored.

The difference between these two values is often one of the most relevant elements during GMP audits, because it allows the real effectiveness of the metrological control system adopted by the laboratory to be assessed.

How to Define Calibration Frequency

One of the most common errors is setting calibration frequency exclusively on a time basis. In reality, the interval should be defined by considering the level of risk associated with the use of the pipette.

The main factors to assess include:

  • the  frequency of instrument use;
  • the criticality of the analysis performed;
  • the history of deviations;
  • the type of samples handled;
  • the results of previous verifications;
  • operating and environmental conditions.

A pipette used daily in critical QC activities will require far more frequent checks than an instrument used only occasionally.

The Role of Preventive Maintenance

Calibration alone is not sufficient to guarantee that performance is maintained over time. Pipettes are mechanical instruments subject to wear: seals, pistons and internal components can deteriorate progressively, altering precision and accuracy without the problem being immediately perceptible to the operator.

Correct preventive maintenance makes it possible to reduce the risk of out-of-tolerance conditions, to increase the stability of measurements and extend the useful life of the instrument. In more structured laboratories, maintenance and calibration are therefore managed as closely integrated activities rather than as separate processes.

This approach reduces the risk of analytical OOS results and maintains GMP compliance over time, improving the overall robustness of the quality system.

Common Errors in Pipette Management

During audits and inspections, recurring issues in pipette management frequently emerge.

Among the most common problems:

  • pipettes used without periodic verification;
  • absence of documentary traceability;
  • failure to evaluate as found data;
  • uncontrolled environmental conditions during verification;
  • failure to manage the impact on analytical results;
  • absence of integration into the overall calibration plan.

Very often the problem is not the absence of calibration, but the lack of a coherent strategy for managing the instrument within the quality system.

Pipette Calibration as Part of the Quality System

In GMP-compliant pharmaceutical laboratories, pipette management must be considered an integral part of the quality system.

This means building a process that includes:

  • verification planning;
  • risk assessment;
  • documentation management;
  • preventive maintenance;
  • periodic trend review;
  • deviation and CAPA management.

The objective is not simply to keep instruments within tolerance, but to guarantee the quality and the defensibility of analytical data over time.